Skip to main navigation Skip to search Skip to main content

False remote monitoring alerts from explanted cardiac implantable electronic device: How is this possible?

Research output: Contribution to journal › Article › peer-review

Abstract

A 72-year-old woman with a dual-chamber implantable cardioverter-defibrillator (Biotronik Lumax 540 DR-T) at elective replacement indicator presented for generator replacement. A new MicroPort generator (Platinium DR) was attached to her existing leads. Eight days later, multiple red alert messages were received on the Biotronik remote monitoring system from the explanted generator. Investigations revealed alert transmission via a CardioMessenger Smart mobile device registered to another patient that came into proximity of the explanted generator. The Biotronik remote monitoring system is unique in that red alerts could be sent through any CardioMessenger Smart device regardless of whether they were paired.

Original languageEnglish
Pages (from-to)132-133
Number of pages2
JournalJournal of Electrocardiology
Volume62
DOIs
Publication statusPublished or Issued - 1 Sept 2020
Externally publishedYes

Keywords

  • Cardiac implantable
  • Electronic device
  • False alerts
  • Remote monitoring

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this