TY - JOUR
T1 - Narrow- and open-angle measurements with anterior-segment optical coherence tomography and Pentacam™
AU - Mou, Dapeng
AU - Fu, Jing
AU - Li, Shuning
AU - Wang, Lan
AU - Wang, Xiaozhen
AU - Wu, Gewei
AU - Qing, Guoping
AU - Peng, Yi
AU - Wang, Ningli
N1 - Copyright:
Copyright 2011 Elsevier B.V., All rights reserved.
PY - 2010/11
Y1 - 2010/11
N2 - ■ BACKGROUND AND OBJECTIVE: To assess the agreement of the Pentacam (Oculus, Wetzlar, Germany) and anterior-segment optical coherence tomography (AS-OCT) in measuring open and narrow angles. ■ PATIENTS AND METHODS: Prospective observational case series in which 39 healthy normal subjects (39 eyes) and 25 patients with narrow angles (37 eyes) were enrolled between May and September 2008. The anterior chamber measurements were performed by AS-OCT and the Pentacam. ■ RESULTS: There was no significant difference in the open-angle measurements between the Pentacam and AS-OCT (P > .05). However, in patients with narrow angles, the average values of nasal and temporal anterior chamber angles taken by the Pentacam were larger than those of AS-OCT (25.5° ± 5.66° and 25.77° ± 5.15° vs 13.40° ± 6.81° and 12.13° ± 6.47°; P < .001). ■ CONCLUSION: Both AS-OCT and the Pentacam can reliably measure anterior chamber angles in healthy normal subjects. The difference in measuring narrow angles by the two instruments was due to inability to view the angle recess or the scleral spur with the Pentacam.
AB - ■ BACKGROUND AND OBJECTIVE: To assess the agreement of the Pentacam (Oculus, Wetzlar, Germany) and anterior-segment optical coherence tomography (AS-OCT) in measuring open and narrow angles. ■ PATIENTS AND METHODS: Prospective observational case series in which 39 healthy normal subjects (39 eyes) and 25 patients with narrow angles (37 eyes) were enrolled between May and September 2008. The anterior chamber measurements were performed by AS-OCT and the Pentacam. ■ RESULTS: There was no significant difference in the open-angle measurements between the Pentacam and AS-OCT (P > .05). However, in patients with narrow angles, the average values of nasal and temporal anterior chamber angles taken by the Pentacam were larger than those of AS-OCT (25.5° ± 5.66° and 25.77° ± 5.15° vs 13.40° ± 6.81° and 12.13° ± 6.47°; P < .001). ■ CONCLUSION: Both AS-OCT and the Pentacam can reliably measure anterior chamber angles in healthy normal subjects. The difference in measuring narrow angles by the two instruments was due to inability to view the angle recess or the scleral spur with the Pentacam.
UR - http://www.scopus.com/inward/record.url?scp=79953235697&partnerID=8YFLogxK
U2 - 10.3928/15428877-20100929-06
DO - 10.3928/15428877-20100929-06
M3 - Article
C2 - 20954643
AN - SCOPUS:79953235697
SN - 1542-8877
VL - 41
SP - 622
EP - 628
JO - Ophthalmic Surgery Lasers and Imaging
JF - Ophthalmic Surgery Lasers and Imaging
IS - 6
ER -