T1 - Trabecular metal acetabular components in primary total hip arthroplasty
T2 - Higher risk for revision compared with other uncemented cup designs in a collaborative register study including 93,709 hips
AU - Laaksonen, Inari
AU - Lorimer, Michelle
AU - Gromov, Kirill
AU - Eskelinen, Antti
AU - Rolfson, Ola
AU - Graves, Stephen E.
AU - Malchau, Henrik
AU - Mohaddes, Maziar
N1 - Publisher Copyright:
© 2018 The Author(s). Published by Taylor & Francis on behalf of the Nordic Orthopedic Federation.
PY - 2018/5/4
Y1 - 2018/5/4
N2 - Background and purpose — Trabecular metal (TM) cups have demonstrated favorable results in acetabular revision and their use in primary total hip arthroplasty (THA) is increasing. Some evidence show that TM cups might decrease periprosthetic infection (PPI) incidence. We compared the survivorship of TM cups with that of other uncemented cups in primary THA, and evaluated whether the use of TM cups is associated with a lower risk of PPI. Patients and methods — 10,113 primary THAs with TM cup and 85,596 THAs with other uncemented cups from 2 high-quality national arthroplasty registries were included. The mean follow-up times were 3.0 years for the TM cups and 3.8 years for the other uncemented cups. Results — The overall survivorship up to 8 years for TM cups and other uncemented cups was 94.4% and 96.2%, respectively (p = < 0.001). Adjusting for relevant covariates in a Cox regression model the TM cups had a persistently higher revision risk than other uncemented cups (HR =1.5, 95% CI 1.4–1.7, p = < 0.001). There was a slightly higher, though not statistically significant, revision rate for PPI in the TM group (1.2, 95% CI 1.0–1.6, p = 0.09). Interpretation — Risk of revision for any reason was higher for the TM cup than for other uncemented cups in primary THA. In contrast to our hypothesis, there was no evidence that the revision rate for PPI was lower in the TM cup patients. Regardless of the promising early and mid-term results for TM cups in hip revision arthroplasty, we would like to sound a note of caution on the increasing use of the TM design, especially in uncomplicated primary THAs, where uncemented titanium cups are considered to provide a reliable outcome.
AB - Background and purpose — Trabecular metal (TM) cups have demonstrated favorable results in acetabular revision and their use in primary total hip arthroplasty (THA) is increasing. Some evidence show that TM cups might decrease periprosthetic infection (PPI) incidence. We compared the survivorship of TM cups with that of other uncemented cups in primary THA, and evaluated whether the use of TM cups is associated with a lower risk of PPI. Patients and methods — 10,113 primary THAs with TM cup and 85,596 THAs with other uncemented cups from 2 high-quality national arthroplasty registries were included. The mean follow-up times were 3.0 years for the TM cups and 3.8 years for the other uncemented cups. Results — The overall survivorship up to 8 years for TM cups and other uncemented cups was 94.4% and 96.2%, respectively (p = < 0.001). Adjusting for relevant covariates in a Cox regression model the TM cups had a persistently higher revision risk than other uncemented cups (HR =1.5, 95% CI 1.4–1.7, p = < 0.001). There was a slightly higher, though not statistically significant, revision rate for PPI in the TM group (1.2, 95% CI 1.0–1.6, p = 0.09). Interpretation — Risk of revision for any reason was higher for the TM cup than for other uncemented cups in primary THA. In contrast to our hypothesis, there was no evidence that the revision rate for PPI was lower in the TM cup patients. Regardless of the promising early and mid-term results for TM cups in hip revision arthroplasty, we would like to sound a note of caution on the increasing use of the TM design, especially in uncomplicated primary THAs, where uncemented titanium cups are considered to provide a reliable outcome.
UR - http://www.scopus.com/inward/record.url?scp=85041557690&partnerID=8YFLogxK
U2 - 10.1080/17453674.2018.1431445
DO - 10.1080/17453674.2018.1431445
M3 - Article
C2 - 29400118
AN - SCOPUS:85041557690
SN - 1745-3674
VL - 89
SP - 259
EP - 264
JO - Acta Orthopaedica
JF - Acta Orthopaedica
IS - 3
ER -